Flow Pulsation in the BP7: Effects of Drive Frequency and a Pulsation Damper
The BP7’s measured flow signal changes with its operating conditions and the use of a pulsation damper. In the measurements presented here, a damper visibly reduces periodic flow fluctuations at a drive frequency of 16 Hz. At 800 Hz, the recorded signals appear steadier—but the measurement does not establish the full amplitude of rapid flow pulsation.
For developers, the practical question is therefore twofold: how steady does the flow need to be, and can the measurement system resolve the fluctuations that matter to the application?
Why a piezoelectric diaphragm pump produces pulsation
The Bartels Pump | BP7 moves fluid through cyclic diaphragm motion. Its piezoelectric actuators repeatedly change the volume of the pump chambers. This pumping action can produce a time-varying flow, even when the average delivery rate remains stable.
Drive frequency and voltage amplitude influence pump operation. The measured flow also reflects the connected fluidic system and the sensor used to observe it. For background on actuation, see Bartels’ article on driving the BP7. Driving the BP7 Micropump: Signal Control and Electronics.
What was measured
The experiment used a BP7 pumping water at room temperature, driven by an mp-Multiboard2 with an mp-Lowdriver. Flow was measured using a Sensirion SLF3S-0600F at a reported acquisition rate of 1,000 samples per second. According to the experiment owner, no additional filtering was applied.
Four operating points were recorded, each with and without a pulsation damper:
| Position in the figure | Drive frequency | Drive voltage |
|---|---|---|
| Upper left | 800 Hz | 55 Vpp |
| Upper right | 800 Hz | 161 Vpp |
| Lower left | 16 Hz | 16 Vpp |
| Lower right | 16 Hz | 75 Vpp |
Each panel shows a 400 ms interval. Blue traces represent measurements with a damper; orange traces represent measurements without one.
The two flow ranges were described as nominal targets of 50 and 500 µl/min. The plotted readings differ from those targets, so the figure should be interpreted using its measured axes rather than as a demonstration of target-flow accuracy.
At 16 Hz, the damper reduces visible fluctuations
Both lower panels show pronounced periodic variations without a damper. With the damper, the recorded flow varies over a visibly narrower range.
This supports a qualitative conclusion for the tested setup: the damper reduces the flow fluctuations visible to the sensor at these low-frequency operating points. It does not eliminate all variation.
The mean readings also differ between the damped and undamped traces, particularly in the higher-flow panel. These records therefore compare the systems at the stated drive settings; they are not comparisons at exactly matched measured mean flow.
At 800 Hz, a steady-looking trace needs careful interpretation
The upper panels show smaller recorded fluctuations than the lower panels. However, acquisition rate and measurement bandwidth must be considered before interpreting this as a reduction in actual high-frequency pulsation.
At 1,000 samples per second, the Nyquist frequency is 500 Hz. An 800 Hz flow component cannot be uniquely reconstructed from that recording using conventional sampling. Faster components may alias into the recorded frequency range if they reach the acquisition stage. NI: Bandwidth, Nyquist Sampling Theorem, and Aliasing.
Sensor dynamics impose a separate limitation. Sensirion specifies a response time below 20 ms for the SLF3S-0600F. That specification does not demonstrate faithful amplitude measurement at 800 Hz, whose cycle lasts only 1.25 ms. This is a measurement limitation inferred from the sensor specification and the experiment’s timing. SLF3S-0600F datasheet, version 1.2, October 2023.
The defensible conclusion is that the recorded sensor output appears steadier at 800 Hz. The figure alone cannot quantify the full instantaneous flow ripple at that operating frequency.
How to use these results in system development
Start by defining the relevant requirement: stable average delivery, limited short-term flow variation, or both. Then evaluate the pump, damper and sensor together under the intended operating conditions.
For a meaningful comparison:
- Compare damped and undamped operation at matched measured mean flow where possible.
- Record the fluidic layout, tubing, backpressure and damper position.
- Choose sensor bandwidth and acquisition settings suited to the fluctuations of interest.
- Define the pulsation metric explicitly, including its reference value and measurement interval.
The percentages annotated in the supplied figure are not used here as quantitative specifications because their calculation method has not yet been verified.
Applying the findings with FluidCube
The measurements were acquired using the earlier Multiboard2 platform. For current development work, FluidCube supports modular BP7 driving electronics, including the mp-Lowdriver. Setup information is available in the FluidCube and FluidicStudio manual. Driving the BP7 Micropump: Signal Control and Electronics; FluidCube and FluidicStudio Manual.
The historical traces remain evidence for the tested configuration. Reproducing their behavior with FluidCube requires checking the drive parameters and fluidic conditions.
The clearest result of this experiment is the visible reduction in measured flow fluctuations with a damper at 16 Hz. At higher drive frequencies, reliable pulsation assessment requires a measurement chain capable of resolving the relevant dynamics.